Model  | WRSP-602  | 
Syringe Size  | 10, 20, 30, 50/60 ml  | 
Applicable Syringe  | Compatible with syringe of any standard  | 
VTBI  | 0.1-9999 ml        < 1000 ml in 0.1 ml increments                   ≧ 1000 ml in 1 ml increments  | 
Flow Rate  | Syringe 10 ml: 0.1-400 ml/h Syringe 20 ml: 0.1-600 ml/h   <100 ml/h in 0.1 ml/h increments Syringe 30 ml: 0.1-900 ml/h   ≧ 100 ml/h in 1 ml/h increments Syringe 50/60 ml: 0.1-1300 ml/h  | 
Bolus Rate  | Syringe 10 ml: 400 ml/h Syringe 20 ml: 600 ml/h    Syringe 30 ml: 900 ml/h    Syringe 50/60 ml: 1300 ml/h  | 
Anti-Bolus  | Automatic  | 
Accuracy  | ±2% (mechanical accuracy ≦ 1%)  | 
Infusion Mode  | Flow rate: ml/min, ml/h Time-based Body weight: mg/kg/min, mg/kg/h, ug/kg/min, ug/kg/h etc.  | 
KVO Rate  | 0.1-1 ml/h (in 0.1 ml/h increments)  | 
Alarms  | Occlusion, near empty, end program, low battery, end battery, AC   power off, motor malfunction, system malfunction, reminder alarm, pressure   sensor error, syringe installation error, syringe drop off  | 
Additional Features  | Real-time infused volume, automatic power switching, automatic   syringe identification, mute key, purge, bolus, anti-bolus, system memory  | 
Drug Library  | Available  | 
Occlusion Sensitivity  | High, medium, low  | 
Power Supply, AC  | 110/230V (optional), 50-60 Hz, 20VA  | 
Battery  | 9.6±1.6V, rechargeable  | 
Battery Life  | 7 hours at 5 ml/h   | 
Working Temperature  | 5-40℃  | 
Relative Humidity  | 20-90%  | 
Atmospheric Pressure  | 860-1060hpa  | 
Size  | 314*167*140 mm  | 
Weight  | 2.5kg  | 
Safety Classification  | Class Ⅱ, type CF  |