Feature:
1.Display mode High-definition 10.4” TFT color LCD screen display
2.Ventilation mode IPPV, SIPPV, SIMV, PEEP, MANUAL, SIGH, IMV, VCV, PCV
3.Ventilation parameters Tidal volume 20~2000 ml
Rate 1~120 bpm
SIMV rate 2~20 bpm
I:E 4:1~1:8
Pressure trigger sensitivity -1.0-2.0kPa
Flow trigger sensitivity 1-20L/min
PEEP 0~2.0Kpa
Pressure range 3~60cmH2O
Pressure support 3~60cmH2O
SIGH 1.5 times the inspiratory time(60-120 adjustable)
Inspiratory platform 0~50%
Oxygen concentration 21~100%
Pressure limitation 5~60cmH2O
Pressure slope 1~10gear
Flow rate setting 1~10gear
Expiratory trigger pressure -1.0~2.0Kpa
Time control 1~10s
4.Parameters for ventilation monitoring: tidal volume, ventilation volume, IPPV rate, SIMV rate, total respiratory rate, I/E,
peak pressure of airway, average pressure, pressure - time waveform, flow rate - time waveform, PEEP, inspiratory trigger
pressure, Inspiratory platform (autonomous inspiratory rate, lung compliance, airway resistance, flow -volume loop, pressure
-volume loop)
5.Monitoring of oxygen concentration: 21%~100%
6.Security Alarm System
1)Oxygen Concentration alarm upper limit setting range 21%~100%, low limit setting range 10%~80%
2)Airway pressure alarm upper limit setting range 0.3~6.0 kPa,low limit setting range 0.2~5.0 kPa
3)Per-minute ventilation volume alarm upper limit setting range 3.0~30L/min,low limit setting range 1.0~10L/min
4)Respiratory rate upper limit setting range 50~99bpm, low limit setting range 0~50bpm
5)Tidal volume upper limit setting range 10~2000ml, low limit setting range 0~1800ml
6)Sustained high-pressure alarm it will give alarm when stress have consistently been higher than 2.5 kPa
7)Suffocation alarm it will give sound and light alarm if there is no tidal volume input for 15 seconds
8)Power alarm
9) Gas shortage alarm
10) Intubation off alarm
7. Power: AC 220V±10% 50Hz±1Hz , UPS, Storage battery
8.Mechanical Arms
9.(Air compressor)
*Items within brackets are optional.